Measurement of primary haemostasis using a pressure clamp technique
V Kretschmer1, S Bade, M Weippert-Kretschmer
1Department of Transfusion Medicine and Haemostaseology, University Hospital, Marburg, Germany. kretschv@mailer.uni-marburg.de
Unlabelled:
With the PFA-100 a sensitive and specific screening test for primary haemostasis has recently become available. An important part of the device is a capillary, providing a defined haemodynamic resistance for the perfusion of the aperture. A modified method to measure platelet function (VCP2) is presented in which the capillary essentially is replaced with an 'electronic capillary' by clamping the pressure/flow relationship.
Results And Conclusion:
Closure time (CT) and blood volume (BV) as determined by PFA-100 and VCP2 correlated well within (r = 0.922 - 0.952) and between the two methods (r = 0.86). The test variability (CV) of CT could be significantly reduced in the VCP2 method (collagen/epi 3.9 vs. 5.9%, p<0.05; collagen/ADP 3.3 vs. 6.9%, p<0.001), thus considerably increasing test reliability and reducing test variance. In preliminary clinical studies the VCP2 system showed comparable sensitivity for vWD and slightly less sensitivity regarding ASA ingestion. The test spectrum of VCP2 could be extended to more thrombocytopenic samples (< or =20 000/microl) even in combination with low haematocrit levels (20%), thus perhaps permitting the determination of the bleeding risk in bone marrow hypoplasia. Additionally, the sensitivity and applicability can easily be adapted to the desired need only by software modifications.
More Related Videos
Related Concept Videos
Patch Clamp
In this method, a glass micropipette containing electrolyte solution is tightly sealed against a small portion of the cell membrane. As a result, a patch of the cell...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure in the Leg
Preparation:
Measurement of Blood Pressure