Implantable cardioverter defibrillator T wave oversensing caused by hyperglycemia
A Krishen1, R K Shepard, J A Leffler
1Division of Cardiac Electrophysiology, Medical College of Virginia, Virginia Commonwealth University, Richmond 23298-0053, USA.
Abstract:
This case report highlights a previously unreported cause of T wave oversensing in a patient with an ICD and recent onset diabetes. Consistent T wave oversensing was observed at elevated serum glucose levels and this finding was reproduced with a glucose challenge. No T wave oversensing was seen during treatment of hyperglycemia. Alterations in serum chemistry may account for intermittent T wave oversensing in patients with ICDs.