Assessment of pacemaker chronotropic response: implementation of the Wilkoff mathematical model
R A Freedman1, D L Hopper, J Mah
1Department of Medicine, University of Utah, and Veterans Affairs Medical Center, Salt Lake City, USA. roger.freedman@hsc.utah.edu
Abstract:
There are few practical guidelines for proper adjustment of rate responsive pacemaker sensor parameters. This study describes the application of the chronotropic assessment exercise protocol (CAEP) and the Wilkoff model of chronotropic response to assess the adjustment of pacemaker sensor parameters. In 31 patients implanted 1 month previously with a dual sensor pacemaker, pacemaker sensor parameters were adjusted to yield a peak sensor rate of 100 beats/min on a simple 6-minute walk (low intensity treadmill exercise [LITE] protocol); the maximum sensor rate was set to the age predicted maximum heart rate (220-age). The rate response behavior of the pacemaker was then assessed using the slope of metabolic-chronotropic relation (MCR) during CAEP exercise. After adjustments based on the LITE protocol, CAEP exercise yielded MCR slopes of 0.92 +/- 0.25 for the entire study group, which compares well with the predicted normal slope of 1. However, 7 of the 31 patients had sensor MCR slopes during CAEP exercise that were 2 SD or more below expected. To test the sensitivity of this approach to suboptimal pacemaker programming or suboptimal exercise, simulations were performed with the maximum sensor rate programmed below age-predicted maximum heart rate or with exercise truncated before maximum exertion; with these conditions, MCR slopes were sharply lower for the entire group. The authors conclude that a simple treadmill walk (LITE) allowed for optimum programming of sensor parameters in most patients, but in a minority the chronotropic behavior was underresponsive. Failure to appropriately adjust pacemaker maximum sensor rate or failure to achieve peak exercise sharply limits the accuracy of this methodology.
Related Concept Videos
Electrophysiology of Normal Cardiac Rhythm
Mechanism of Cardiac Arrhythmias
RC Circuit with Source
Due to the inherent properties of a capacitor, its voltage cannot change instantaneously. This means that immediately after the switch is closed, the capacitor's voltage remains the same as it was just before the switch was closed.
By...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Pharmacodynamic Models: Direct Effect Model and Indirect Response Model
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials