Tilt Testing with Combined Lower Body Negative Pressure: a "Gold Standard" for Measuring Orthostatic Tolerance
Published on: March 21, 2013
The maintenance of a stable baseline in intra-uterine pressure with varying maternal position--a practical approach
Abstract:
When measuring intra-uterine pressure, the tracing baseline varies according to maternal position because of changes in hydrostatic pressure. To compensate for this effect, the membrane of the transducer should always be kept close to the uterine fundus. An additional advantage of this method is that it enables the use of a short catheter, thus causing less tracing artifacts.