MRI volume measurements of hypointense objects. A phantom study using stereological methods.

Paula Gadeberg1, Hans J G Gundersen, Finn Taagehøj

  • 1Department of Neurology, Aarhus University Hospital, Aarhus University, Aarhus, Denmark. paula@akhphd.au.dk

Journal of Neuroscience Methods
|February 22, 2002
PubMed
Summary

Magnetic Resonance Imaging (MRI) volume estimates for hypointense objects are biased due to slice thickness and cross-talk. A simple correction is not feasible, and measuring maximal contrast is an imperfect solution.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

A Practical Guide to Management of Chronic Inflammatory Demyelinating Polyradiculoneuropathy.

Neurology and therapy·2026
Same author

Predictors of relapses in patients with chronic inflammatory demyelinating polyneuropathy receiving subcutaneous immunoglobulin therapy - a post-hoc analysis.

Journal of neuromuscular diseases·2026
Same author

Prevention of axonal loss after immediate dosage titration of immunoglobulin in multifocal motor neuropathy.

European journal of neurology·2024
Same author

Outcome in chronic inflammatory demyelinating polyneuropathy: A systematic review and meta-analysis.

Muscle & nerve·2023
Same author

Axonal loss at time of diagnosis as biomarker for long-term disability in chronic inflammatory demyelinating polyneuropathy.

Muscle & nerve·2022
Same author

Long-term follow-up of facilitated subcutaneous immunoglobulin therapy in multifocal motor neuropathy.

Journal of the neurological sciences·2021

Area of Science:

  • Medical Imaging
  • Biophysics
  • Stereology

Background:

  • Accurate volume estimation is crucial in medical imaging.
  • Thick MRI slices introduce bias in volume measurements, particularly for hypointense (dark) objects.
  • Existing correction models are insufficient for hypointense objects.

Purpose of the Study:

  • To quantify the bias in MRI volume estimation of hypointense objects.
  • To investigate the impact of slice thickness and contrast concentration on measurement bias.
  • To explore potential correction methods for MRI-derived hypointense object volumes.

Main Methods:

  • A stereological MRI protocol was used with phantoms of known hypointense object diameters.
  • Measurements were taken using 1, 3, and 5 mm slices with varying contrast concentrations.

Related Experiment Videos

  • Bias was estimated by comparing MRI-derived volumes to true object sizes.
  • Main Results:

    • MRI volume estimates underestimated the true size of hypointense objects.
    • Interslice cross-talk, where an object appears in adjacent slices, significantly contributed to the bias.
    • A simple geometric correction was not possible due to cross-talk; an empirical correction for phantoms had unpredictable validity for biological tissues.

    Conclusions:

    • MRI measurements of hypointense objects are subject to significant bias, primarily due to interslice cross-talk.
    • Developing a universally applicable correction for this bias remains challenging.
    • Measuring maximal contrast from edge to edge is a proposed, though imperfect, workaround.