Related Experiment Videos
Amperometric quantification of total coliforms and specific detection of Escherichia coli
Adrian Schwartz Mittelmann1, Eliora Z Ron, Judith Rishpon
1Department of Molecular Microbiology and Biotechnology, Tel-Aviv University, Ramat-Aviv, Israel.
Abstract:
The quantitative determination of total and fecal coliforms, as indicators of fecal pollution, is essential for water quality control. We developed a sensitive, inexpensive amperometric enzyme biosensor based on the electrochemical detection of beta-galactosidase activity, using p-amino-phenyl-beta-D-galactopyranoside as substrate, for determining the density of coliforms, represented by Escherichia coli and Klebsiella pneumoniae. The specific detection of E. coli was achieved using an antibody-coated electrode that specifically binds the target bacteria. Amperometric detection enabled the determination of 1000 colony-forming units/mL within 60-75 min. Preincubation for 5-6 h further increased the sensitivity more than 100-fold. The present experimental setup allowed the simultaneous analysis of up to eight samples, using disposable screen-printed electrodes.