Model-based administration of inhalation anaesthesia. 4. Applying the system model
Related Experiment Video
Updated: Aug 10, 2026

Use of a Low-flow Digital Anesthesia System for Mice and Rats
Published on: September 7, 2016
J G C Lerou1, R Verheijen, L H D J Booij
1Institute for Anaesthesiology, University of Nijmegen, The Netherlands.
Background:
We developed and tested a simple dosing strategy for rapid induction with isoflurane followed by maintenance under minimal-flow conditions, that is 0.5 litre min(-1) total fresh gas flow (FGF). An end-expired concentration was to be achieved within 5 min in a desired therapeutic window, that is 0.8-1.1 vol%, and to be maintained within it for at least 30 min.
Methods:
With our new model we computed a three-stage regimen using one fixed vaporizer setting: 3 vol% isoflurane in a FGF of 3 and 1.5 litre min(-1), each for 3 min, and 0.5 litre min thereafter. The ratio of nitrous oxide:oxygen was, consecutively, 2:1, 2:1, and 2:3. We evaluated this scheme in 58 adult patients (body mass 74 (SD 13) kg), mostly during eye and ear, nose, and throat surgery.
Results:
Measured oxygen (33-45 vol%) and nitrous oxide concentrations (66-50 vol%) evolved in accordance with those computed. In five patients with a median of body mass 92 kg (range 76-126 kg), inspired oxygen concentrations decreased to less than 30 vol%. End-expired isoflurane concentration entered the window after 2 min (range 1.0-5.67 min) and attained its maximum, that is 0.96 vol% (0.8-1.2 vol%), after 3.45 min (1.67-6.33 min). The mean end-expired concentration was in the desired window from 3-60 min and an average of 72% of individual measurements were within the window from 3-30 min. The scheme was adapted in six patients (excluded from analysis) because of hypotension.
Conclusion:
The regimen is easily remembered, reliable, and lends itself to alternative strategies, but must be guided by the monitoring of gas and vapour concentrations and haemodynamic variables.
More Related Videos
14:52Recording Brain Electromagnetic Activity During the Administration of the Gaseous Anesthetic Agents Xenon and Nitrous Oxide in Healthy Volunteers
Published on: January 13, 2018
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Related Concept Videos
Stages of General Anesthesia
Inhalational Anesthetics: Overview
Asthma-IV: Nursing Management
First, in...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant, half-life,...
Application of Integration: Problem Solving