Purpose:
This study was conducted to evaluate patients' proficiency in self-monitoring of blood glucose (SMBG).
Methods:
Diabetes nurse educators in 4 suburban Minneapolis clinic sites surveyed the SMBG training/cure practices of 280 patients with type 1 and type 2 diabetes. Participant SMBG technique was measured by direct observation. Participants performed a finger puncture and used their own meters to measure the first blood sample. A second sample was measured on the HemoCue B Glucose analyzer, and a third sample was used to measure hemoglobin. The series of tests were then repeated. If either of the 2 glucose tests was more than 15% from the HemoCue value, participants were reeducated about the manufacturer's suggested procedure.
Results:
Of the 280 participants, 19% had blood glucose test results greater than the 15% limit for meter accuracy. After reeducation, 69% of those who had initially failed achieved acceptable results. The most significant problems were lack of periodic meter technique evaluation, difficulty using wipe meters, incorrect use of control solutions, lack of hand washing even when observed, and unclean meters.
Conclusions:
As a result of the study, guidelines were subsequently developed to evaluate meter accuracy in an outpatient setting. Further effort is needed to establish standards for evaluating SMBG.
More Related Videos
08:01A Method for Manipulating Blood Glucose and Measuring Resulting Changes in Cognitive Accessibility of Target Stimuli
Published on: August 12, 2016
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Strategies of Self-Presentation III: Self-Monitoring
Type II Diabetes Mellitus III: Clinical Manifestations and Diagnosis
Hyperglycemia