Evaluation of the new Ocuton S tonometer
Giorgio Marchini1, Silvia Babighian, Luca Specchia
1Department of Ophthalmology,University of Verona, Italy. oculist@borgotrento.univr.it
Acta Ophthalmologica Scandinavica
|April 16, 2002
Summary
The Ocuton S tonometer is safe and easy for patient self-tonometry, but it overestimates intraocular pressure (IOP) compared to Goldmann tonometry, requiring further improvements for clinical reliability.
Area of Science:
- Ophthalmology
- Glaucoma diagnostics
- Intraocular pressure measurement
Background:
- Accurate intraocular pressure (IOP) measurement is crucial for diagnosing and managing glaucoma.
- The Ocuton S tonometer offers a potential alternative for IOP assessment.
- Evaluating new tonometry devices against established standards like Goldmann tonometry is essential.
Purpose of the Study:
- To assess the Ocuton S tonometer's intra- and interobserver variability.
- To compare Ocuton S tonometer readings with Goldmann tonometry.
- To evaluate the reliability of patient self-tonometry using the Ocuton S and its safety.
Main Methods:
- The study included 35 healthy subjects and 45 patients with primary open-angle glaucoma (POAG).
- Tonometry was performed using the Ocuton S and Goldmann tonometers, with central ultrasonic pachymetry, keratometry, and biomicroscopy also evaluated.
- Intra- and interobserver variability and self-tonometry reliability were assessed in subgroups.
Main Results:
- Intra- and interobserver variability for the Ocuton S ranged from 0.38 to 0.66.
- Ocuton S consistently overestimated IOP compared to Goldmann tonometry (24.4 mmHg vs 18.1 mmHg, p < 0.0001).
- A significant correlation was found between corneal thickness and IOP for both devices (r ≈ 0.52, p < 0.02). Self-tonometry showed no significant difference from expert measurements.
Conclusions:
- The Ocuton S tonometer is safe and can be used by patients for self-measurement.
- The device overestimates IOP relative to the Goldmann tonometer.
- Further technical and methodological refinements are needed to enhance the Ocuton S tonometer's clinical reliability.