Direct calibration of a reference standard against the air kerma strength primary standard, at 192Ir HDR energy
K N Govinda Rajan1, T Palani Selvam, B C Bhatt
1Radiological Physics & Advisory Division, BARC, Mumbai, India. grajanmpss@vsnl.com
Abstract:
The primary standard of low air kerma rate sources or beams, maintained at the Radiological Standards Laboratory (RSL) of the Bhabha Atomic Research Centre (BARC), is a 60 cm3 spherical graphite ionization chamber. A 192Ir HDR source was standardized at the hospital site in units of air kerma strength (AKS) using this primary standard. A 400 cm3 bakelite chamber, functioning as a reference standard at the RSL for a long period, at low air kerma rates (compared to external beam dose rates), was calibrated against the primary standard. It was seen that the primary standard and the reference standard, both being of low Z, showed roughly the same scatter response and yielded the same calibration factor for the 400 cm3 reference chamber, with or without room scatter. However, any likelihood of change in the reference chamber calibration factor would necessitate the re-transport of the primary standard to the hospital site for re-calibration. Frequent transport of the primary standard can affect the long-term stability of the primary standard, due to its movement or other extraneous causes. The calibration of the reference standard against the primary standard at the RSL, for an industrial type 192Ir source maintained at the laboratory, showed excellent agreement with the hospital calibration, making it possible to check the reference chamber calibration at RSL itself. Further calibration procedures have been developed to offer traceable calibration of the hospital well ionization chambers.
More Related Videos
09:18Laser-heating and Radiance Spectrometry for the Study of Nuclear Materials in Conditions Simulating a Nuclear Power Plant Accident
Published on: December 14, 2017
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Flame Photometry: Lab
IR Spectrometers
Standard Solutions
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...