Related Experiment Videos
Three-dimensional magnetic resonance coronary angiography using a new blood pool contrast agent: initial experience
Willemijn L F Bedaux1, Mark B M Hofman, Piotr A Wielopolski
1Department of Cardiology, Vrije Universiteit Medical Centre, Amsterdam, The Netherlands. wlf.bedaux@vumc.nl
Objective:
The aim of this feasibility study was to assess the effect of a new blood pool contrast agent on magnetic resonance coronary angiography (MRCA) in patients suspected of having coronary artery disease.
Methods:
Nine patients referred for diagnostic x-ray coronary angiography in the evaluation of chest pain underwent MRCA using a thin-slab three-dimensional (3D) breath-hold segmented gradient echo technique at 1.5 T before and after intravenous administration of feruglose, a new blood pool contrast agent.
Results:
A total of 36 coronary arteries, including 15 stenoses were evaluated. Overall agreement of visually assessed stenoses > 50% was 86%. The mean signal-to-noise ratio in the left anterior descending artery (LAD) did not change significantly after contrast administration, a relative improvement of 1.2 +/- 0.4 (p = 0.1), whereas the mean contrast-to-noise ratio of blood to myocardium improved by 3 +/- 2 (p = 0.01).
Conclusion:
A thin-slab 3D breath-hold scan combined with feruglose allows rapid localization and coverage of the proximal coronary arteries.