Clutter filters adapted to tissue motion in ultrasound color flow imaging
Steinar Bjaerum1, Hans Torp, Kjell Kristoffersen
1GE Vingmed Ultrasound, Horten, Norway. steinar.bjaerum@med.ge.com
Abstract:
The quality of ultrasound color flow images is highly dependent on sufficient attenuation of the clutter signals originating from stationary and slowly moving tissue. Without sufficient clutter rejection, the detection of low velocity blood flow will be poor, and the velocity estimates will have a large bias. In some situations, e.g., when imaging the coronary arteries or when the operator moves the probe in search for small vessels, there is considerable movement of tissue. It has been suggested that clutter rejection can be improved by mixing down the signal with an estimate of the mean frequency prior to high pass filtering. In this paper, we compare this algorithm with several other adaptive clutter filtering algorithms using both experimental data and simulations. We found that realistic accelerations of the tissue have a large effect on the clutter rejection. The best results were obtained by mixing down the signal with non-constant phase increments estimated from the signal. This adapted the filter to a possibly accelerated tissue motion and produced a significant improvement in clutter rejection.