Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method
Published on: May 15, 2009
Practice audits: reliability of sphygmomanometers and blood pressure recording bias
1Birmingham South West Primary Care Group, Norton Court, K Block, Birmingham Womens Hospital, Metchley Park Road, Edgbaston, UK. shaukat.ali@pc.birminghamha.wmids.nhs.uk
Abstract:
It is well established that numerous errors, biases and omissions in recording blood pressure exist. This study had two objectives. Firstly, to measure the accuracy of sphygmomanometers used in general practice and secondly to assess digit bias in blood pressure recording. This study was carried out in the then Northfield PCG, which comprised 18 practices and 67 GPs. A total of 131 mercury and aneroid sphygmomanometers were tested for accuracy by a trained technician in accordance with the methods specified in BS 2743 (1990). Accuracy was defined as an error of greater than 10 mm Hg. The second part of the methodology involved undertaking an audit of the proportion of registered patients aged 35-80 years who had their blood pressure measured within the last 5 years by members of the Primary Health Care Team. The results were that of the mercury and aneroid sphygmomanometers tested, 17% were inaccurate. Of these, 4% recorded an error greater than 10 mm Hg. One percent of mercury and 10% of aneroid sphygmomanometers recorded an error greater than 10 mm Hg respectively. Sixteen (12%) sphygmomanometers were so deteriorated (air leaks, dirt in mercury) that the researcher suggested their immediate withdrawal from service. The results of the blood pressure recording audit suggested digit bias of both systolic and diastolic recordings to the nearest 10 mm Hg. This study suggests that sphygmomanometers used in general practice are very likely to be inaccurate and some may well be so deteriorated that they should be withdrawn from service. The results of the blood pressure audit showed digit bias in systolic and diastolic readings to the nearest 10 mm Hg. The implications for clinical care-both over diagnosis and under diagnosis-although not assessed are likely to be appreciable. PCG Clinical Governance teams in conjunction with Practice Clinical Leads must address these basic issues.
More Related Videos
14:28Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
Published on: June 27, 2025
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure