Related Experiment Videos
Determination of sulpiride by capillary electrophoresis with end-column electrogenerated chemiluminescence detection
Jifeng Liu1, Weidong Cao, Haibo Qiu
1State Key Laboratory of Electroanalytical Chemistry, Changchun Institute of Applied Chemistry, Chinese Academy of Sciences, Changchun, Jilin 130022, People's Republic of China.
Background:
Capillary electrophoresis (CE) with tris(2,2'-bipyridyl)ruthenium(II) [Ru(bpy)3(2+)]-electrogenerated chemiluminescence (ECL) detection is a promising method for clinical analysis. In this study, a method combining CE with Ru(bpy)3(2+) ECL (CE-ECL) detection that can be applied to amine-containing clinical species was developed, and the performance of CE-ECL as a quantitative method for determination of sulpiride in human plasma or urine was evaluated.
Methods:
Sulpiride was separated by capillary zone electrophoresis in uncoated fused-silica capillaries [50 cm x 25 microm (i.d.)] filled with phosphate buffer (pH 8.0) and a driving voltage of +15 kV, with end-column Ru(bpy)3(2+) ECL detection. A platinum disc electrode was used as working electrode. Sulpiride in human plasma or urine samples (100 microL) was extracted by a double-step liquid-liquid extraction procedure, dried under nitrogen at 35 degrees C in a water bath, and reconstituted with 100 microL of filtered water. The extraction solvent was ethyl acetate-dichloromethane (5:1 by volume).
Results:
Under optimum conditions (pH 8.0 phosphate buffer, injection for 6 s at 10 kV, and +1.2 V as detection potential), separation of sulpiride was accomplished within 4 min. The calibration curve was linear over a concentration range of 0.05-25.0 micromol/L, and the limit of detection was 2.9 x 10(-8) mol/L for sulpiride. Intra- and interday CVs for ECL intensities were <6%. Extraction recoveries of sulpiride were 95.6-101% with CVs of 2.9-6.0%. The method was clinically validated for patient plasma and urine samples.
Conclusions:
CE combined with Ru(bpy)3(2+) ECL is reproducible, precise, selective, and enables the analysis of sulpiride in human plasma and urine. It thus is of value for rapid and efficient analysis of amine-containing analytes of clinical interest.