Prothrombin time quality assessment scheme in the Basque Country
M Vacas1, P J Lafuente, M J Aguirrebeitia
1FIDEC-Fundación para la Investigación y Docencia de las Enfermedades Cardiovasculares, Bilbao, Spain. bdbvarim@lg.ehu.es
Thrombosis Research
|July 2, 2002
Summary
The Basque Country Quality Assessment Scheme on Oral Anticoagulant Treatment (PEEC-CAV) established quality controls for prothrombin time (PT) testing. The study evaluated Thromboplastin Bilbao (TBi) reagent performance, showing consistent results over time.
Area of Science:
- Clinical Chemistry
- Hematology
- Quality Control in Laboratory Medicine
Background:
- The Basque Country Quality Assessment Scheme on Oral Anticoagulant Treatment (PEEC-CAV) was initiated in 1984.
- The scheme integrated prothrombin time (PT) quality controls with a novel in-house standard reagent, Thromboplastin Bilbao (TBi).
- TBi transitioned from human to rabbit brain origin in 1994, aligning with WHO guidelines.
Purpose of the Study:
- To assess diverse PT testing system performances across hospitals in the Basque Country.
- To evaluate the responsiveness and consistency of the TBi reagent within quality control procedures.
- To compare the performance of TBi with previous human-origin thromboplastin reagents.
Main Methods:
- Nine hospitals participated in quality control assessments.
- Prothrombin time (PT) testing systems, including routine laboratory reagents and coagulometers, were evaluated.
- The Thromboplastin Bilbao (TBi) reagent was tested using both routine and manual techniques.
Main Results:
- The INR coefficient of variation (CV) for control plasmas was 8.29+/-4.18% using routine laboratory systems (1994-1999).
- Using TBi with a manual technique, the INR CV was 9.7+/-2.70%.
- These results demonstrated consistency comparable to earlier PT ratio CVs obtained with human brain thromboplastin (9.26+/-2.84% from 1984-1988).
Conclusions:
- The TBi reagent, whether rabbit or human-derived, provided reliable and consistent performance in PT quality controls.
- The PEEC-CAV scheme effectively monitored PT test performance and reagent stability over extended periods.
- The findings support the continued use of standardized reagents like TBi for ensuring accuracy in oral anticoagulant therapy monitoring.