Determination of trace level bromate and perchlorate in drinking water by ion chromatography with an evaporative
Yongjian Liu1, Shifen Mou, Shawn Heberling
1Research Center for Eco-Environmental Sciences, Chinese Academy of Sciences, Beijing.
Abstract:
A simple sample preconcentration technique employing microwave-based evaporation for the determination of trace level bromate and perchlorate in drinking water with ion chromatography is presented. With a hydrophilic anion-exchange column and a sodium hydroxide eluent in linear gradient, bromate and perchlorate can be determined in one injection within 35 min. Prior to ion chromatographic analysis, the drinking water sample was treated with an OnGuard-Ag cartridge to remove the superfluous chloride and concentrated 20-fold using a PTFE beaker in a domestic microwave oven for 15 min. The recoveries of the anions ranged from 94.6% for NO2- to 105.2% for F-. The detection limits for bromate, perchlorate, iodate and chlorate were 0.1, 0.2, 0.1 and 0.2 microg/l, respectively. The developed method is applicable for the quantitation of bromate and perchlorate in drinking water samples.
Related Concept Videos
Precipitation of Ions
The equation that describes the equilibrium between solid calcium carbonate and its solvated ions is:
Precipitation Titration Curve: Analysis
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Precipitation and Co-precipitation
Gravimetry: Inorganic And Organic Precipitating Agents
Ion-Exchange Chromatography