Clinical evaluation of the dynamic observing tonometer.

Andrew J Morgan1, Sarah L Hosking, John F Salmon

  • 1Neurosciences Research Institute, School of Life and Health Sciences, Aston University, Birmingham, England.

Journal of Glaucoma
|August 10, 2002
PubMed
Summary

The Dynamic Observing Tonometer offers continuous intraocular pressure (IOP) monitoring with reasonable accuracy and repeatability, comparable to existing devices. This diagnostic contact lens shows potential for improved glaucoma assessment and management.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Organosolv processing of Sitka spruce sawdust: large scale preparation of native-like lignin and lignin<sup>OX</sup> for valorisation.

Faraday discussions·2025
Same author

Focusing of X-ray free-electron laser pulses using multilayer Laue lenses.

Optics express·2025
Same author

The dynamics of competition and decision-making.

Psychonomic bulletin & review·2024
Same author

Resolving Nonequilibrium Shape Variations among Millions of Gold Nanoparticles.

ACS nano·2024
Same author

Using cognitive modeling to examine the effects of competition on strategy and effort in races and tournaments.

Psychonomic bulletin & review·2022
Same author

Megahertz pulse trains enable multi-hit serial femtosecond crystallography experiments at X-ray free electron lasers.

Nature communications·2022

Area of Science:

  • Ophthalmology
  • Medical Devices
  • Diagnostic Technology

Background:

  • Intraocular pressure (IOP) is a critical factor in glaucoma diagnosis and management.
  • Continuous IOP monitoring offers potential advantages over intermittent measurements.
  • The Dynamic Observing Tonometer (DOT) is a novel diagnostic contact lens designed for continuous IOP measurement.

Purpose of the Study:

  • To evaluate the accuracy of the Dynamic Observing Tonometer (DOT) compared to Goldmann applanation tonometry and pneumatonometry.
  • To assess the repeatability of IOP measurements obtained with the DOT.
  • To determine the device's capability in measuring pulse amplitude.

Main Methods:

  • 40 subjects underwent IOP measurement using Goldmann applanation tonometry.

Related Experiment Videos

  • The DOT and a pneumatonometer were used for continuous IOP and pulse amplitude measurements.
  • Accuracy was assessed via mean bias and 95% limits of agreement; repeatability was evaluated using reliability coefficients.
  • Main Results:

    • DOT showed a mean bias of +2.1 mm Hg (95% limits of agreement: -4.0 to +8.2 mm Hg) versus Goldmann tonometry (r = 0.78).
    • Pulse amplitude measurements demonstrated a mean bias of +0.4 mm Hg (95% limits of agreement: -1.6 to +2.3 mm Hg) versus pneumatonometry (r = 0.78).
    • IOP measurement repeatability yielded a reliability coefficient of 0.91, with a mean difference of 0.4 mm Hg between readings.

    Conclusions:

    • The Dynamic Observing Tonometer exhibits a slight positive bias compared to Goldmann tonometry but demonstrates good correlation and acceptable repeatability.
    • The device accurately measures pulse amplitude, correlating well with pneumatonometry.
    • The DOT's performance suggests its potential as a reliable tool for continuous IOP monitoring in clinical settings.