Continuous Venous-Arterial Doppler Ultrasound During a Preload Challenge
Published on: January 20, 2023
[Detection of air embolism by a re-usable Doppler probe integrated in a central venous line--application in-vivo]
1Medizinische Klinik I, Klinikum Krefeld, Germany. o.volk@freenet.de
Problem:
Air embolism frequently occurs under neurosurgical operations performed in the sitting position. Recently we reported the idea of inserting a re-usable doppler probe into a blind-ending lumen of a central venous catheter (Schregel-Volk-Catheter, SVC). In vitro testing in a circulation model showed a high sensitivity: air bubbles as small as 0.5 microliter were reliably detected at a distance up to 4 cm from the tip of the SVC.
Method:
After approval by the local ethics committee pigs were anaesthetised and a cardiopulmonary bypass was connected. During the bypass period a 12 F SVC (Medex Medical, Germany) with one blind-ending lumen was positioned into the proximal vena cava cranialis or vena cava caudalis. An 8 MHz doppler probe (MTB Basler, Suisse) was introduced into the SVC and connected with the doppler device Multi-Dop T (DWL, Germany). Well defined air bubbles (3-6 microliter) were generated by a bubble generator and injected into the right femoral vein or the left vena jugularis interna.
Results:
All bubbles were reliably detected by the re-usable doppler probe. Embolic events can be documented, counted and quantified by the Multi-Dop T.
Conclusion:
Using SVC's with a blind-ending lumen could improve and simplify the detection of air embolism. We see several advantages (e.g. sensitivity, costs) compared with established methods (precordial doppler, TEE) for detection of air embolism.
Related Concept Videos
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: