Related Experiment Video
Updated: Sep 29, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
An empirical model for independent dose verification of the Gamma Knife treatment planning
Nakorn Phaisangittisakul1, Lijun Ma
1Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore 21201, USA. nphai001@umaryland.edu
Abstract:
A formalism for an independent dose verification of the Gamma Knife treatment planning is developed. It is based on the approximation that isodose distribution for a single shot is in the shape of an ellipsoid in three-dimensional space. The dose profiles for a phantom along each of the three major axes are fitted to a function which contains the terms that represent the contributions from a point source, an extrafocal scattering, and a flat background. The fitting parameters are extracted for all four helmet collimators, at various shot locations, and with different skull shapes. The 33 parameters of a patient's skull shape obtained from the Skull Scaling Instrument measurements are modeled for individual patients. The relative doses for a treatment volume in the form of 31 x 31 x 31 matrix of points are extracted from the treatment planning system, the Leksell Gamma-Plan (LGP). Our model evaluates the relative doses using the same input parameters as in the LGP, which are skull measurement data, shot location, weight, gamma-angle of the head frame, and helmet collimator size. For 29 single-shot cases, the discrepancy of dose at the focus point between the calculation and the LGP is found to be within -1% to 2%. For multi-shot cases, the value and the coordinate of the maximum dose point from the calculation agree within +/-7% and +/-3 mm with the LGP results. In general, the calculated doses agree with the LGP calculations within +/-10% for the off-center locations. Results of calculation with this method for the dimension and location of the 50% isodose line are in good agreement with results from Leksell GammaPlan. Therefore, this method can be served as a useful tool for secondary quality assurance of Gamma Knife treatment plans.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification (ADCI) and Dose Estimation
Published on: September 4, 2017