A simple method for quantifying fentanyl, sufentanil, or morphine in discard syringes from anesthesia procedures
Laurel S T Holth1, Robert Friesen, Carol Green
1Department of Biochemistry, St. Boniface General Hospital, Winnipeg, Monitoba, Canada. laurel.holth@dkml.com
Therapeutic Drug Monitoring
|September 28, 2002
Summary
A new method accurately quantifies fentanyl, sufentanil, and morphine in anesthesia discard syringes. This system helps detect drug diversion and protect medical staff from addiction risks.
Area of Science:
- Analytical Chemistry
- Forensic Toxicology
- Pharmacology
Background:
- Fentanyl, sufentanil, and morphine are potent opioids frequently used in anesthesia.
- Healthcare professionals handling these substances face a significant risk of addiction and diversion.
- Effective methods are needed to monitor drug usage and prevent misuse in clinical settings.
Purpose of the Study:
- To develop and validate a method for quantifying fentanyl, sufentanil, and morphine in discard syringes.
- To establish the sensitivity, linearity, and precision of the assay for these opioids.
- To assess the potential for cross-reactivity with other common operating room drugs.
Main Methods:
- Utilized the BioRad REMEDi HS Drug Profiling System for drug quantification.
- Determined the limit of detection and linearity for fentanyl, sufentanil, and morphine.
- Calculated within-run and between-run coefficients of variation (CV) to assess assay precision.
- Tested for cross-reactivity with other anesthetic agents.
Main Results:
- The REMEDi HS system demonstrated accurate quantification for fentanyl, sufentanil, and morphine in discard syringes.
- Established detection limits: fentanyl (0.1 µg/mL), sufentanil (0.5 µg/mL), and morphine (0.5 µg/mL).
- Assay precision showed low coefficients of variation, with within-run CVs ranging from 0.9% to 11.6% and between-run CVs from 2.4% to 11.3%.
- No cross-reactivity was observed with other commonly used operating room drugs.
Conclusions:
- The BioRad REMEDi HS Drug Profiling System provides a rapid and accurate method for quantifying fentanyl, sufentanil, and morphine.
- This validated assay is suitable for detecting drug diversion from anesthesia discard syringes.
- The method contributes to drug accountability and the prevention of substance abuse among medical personnel.