Related Experiment Video
Updated: Aug 11, 2026

Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
Effect of sterilization on the cutting efficiency of PVD-coated nickel-titanium endodontic instruments
1Department of Operative Dentistry, University of Münster, Münster, Germany. eschaef@uni-muenster.de
Aim:
To investigate alterations in cutting efficiency when conventional and titanium nitride (TiN) coated nickel-titanium (NiTi) K-files were exposed to repeated sterilization using an autoclave.
Methodology:
A total of 96 NiTi K-files (size 35) were randomly divided into two groups (A and B) of 48 instruments each. Whilst the instruments of group B were exposed to physical vapour deposition (PVD) creating a coating of a TiN layer, the files of group A were not coated. The instruments of groups A and B were randomly divided into four subgroups of 12 instruments each. A.1/B.1: Instruments were exposed to five cycles of sterilization. A.2/B.2: Instruments were exposed to 10 cycles of sterilization. A.3/B.3: Instruments were immersed in NaOCl for 30 min, rinsed in water, and exposed to five cycles of sterilization. A.C/B.C: Instruments were not sterilized (controls). The cutting efficiency of all files was determined by means of a computer-driven testing device. Special plastic samples with cylindrical canals were used and the maximum penetration depth of the files into the lumen was assessed.
Results:
The TiN-coated instruments of groups B.1, B.2 and B.3 showed no significant difference in comparison with the penetration depths of the controls (P> 0.05). The uncoated files of groups A.1, A.2 and A.3 displayed significantly lower maximum penetration depths (P< 0.05) when compared to the control files.
Conclusions:
Repeated sterilization under autoclave or exposure to sodium hypochlorite (NaOCl) prior to sterilization did not alter the cutting efficiency of PVD-coated NiTi K-files.
More Related Videos
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...