Related Experiment Video
Updated: Sep 28, 2026

Measuring the Stiffness of Ex Vivo Mouse Aortas Using Atomic Force Microscopy
Published on: October 19, 2016
Microtensile measurements of single trabeculae stiffness in human femur
Fabiano Bini1, Andrea Marinozzi, Franco Marinozzi
1Faculty of Engineering, Department of Mechanics and Aeronautics, Mechanical and Thermal Measurements Laboratory, University of Rome La Sapienza, Via Eudossiana 18, 00184 Rome, Italy.
Abstract:
In this paper, the authors perform microtensile tests of single trabeculae excised from a human femur head. One of the main issues of this work is to establish some experimental procedures for preparing and testing the specimens. The use of a well-characterized microtensile apparatus allows for a low intraspecimen dispersion of the measured stiffness. Tensile/compressive tests were chosen because they appear less sensitive to errors in the cross-sectional area measurements with respect to bending tests. By these considerations, some tensile/compressive tests of plate-like trabecular specimens have been carried out. Typical stiffness values are 74.2+/-0.7Nmm(-1) for tensile tests, and 58.9+/-0.6Nmm(-1) for compressive test. Another compressive test performed on a shorter specimen yielded a stiffness value of 148.3+/-5.3Nmm(-1). The maximum applied load was about 0.5N. Rough measurements of specimens sizes yielded a Young's modulus value ranging from 1.41 to 1.89GPa.

