Related Experiment Video
Updated: Sep 28, 2026

Single-throughput Complementary High-resolution Analytical Techniques for Characterizing Complex Natural Organic Matter Mixtures
Published on: January 7, 2019
Gas chromatographic-mass spectrometric method for quantification of
Andrea L Rezemini1, Jorge M Vaz, Lilian R F Carvalho
1Instituto de Química, Universidade de São Paulo, SP, Brazil.
Abstract:
A method for the determination of 3-chloro-4-(dichloromethyl)-5-hydroxy-2[5H]-furanone (MX), in drinking water by GC-MS with a limit of detection of 3.0 microg/l and a limit of quantification of 7.0 microg/l is presented. Clean-up by SPE and extraction of water samples with dichloromethane were carried out before the preconcentration of MX, which was derivatized directly in the injector of the GC, and the MX trimethylsilyl derivative was identified and quantitatively determined by MS.