Temporomandibular Joint Arthroplasty Using Metal-on-Metal and Acrylic-on-Metal Configurations : Wear In Laboratory
A L Lippincott III1, J M Dowling, D Phil
1Engineering Consulting Services Inc Prior Lake.
Metal-on-metal temporomandibular joint (TMJ) implants exhibit significantly lower wear than acrylic-on-metal designs. This finding is crucial for improving long-term TMJ replacement performance and patient outcomes.
Area of Science:
- Biomaterials Science
- Orthopedic Surgery
- Mechanical Engineering
Background:
- Long-term performance of temporomandibular joint (TMJ) implants is significantly influenced by wear.
- Understanding wear mechanisms is critical for developing durable TMJ arthroplasty solutions.
Purpose of the Study:
- To compare the wear performance of metal-on-metal TMJ implants against acrylic-on-metal and polyethylene-on-metal configurations.
- To analyze wear characteristics of retrieved and laboratory-tested TMJ implants.
Main Methods:
- Optical microscopy and scanning electron microscopy (SEM) were used to examine retrieved and laboratory-tested TMJ implants.
- Volumetric wear was quantified for different implant material pairings under laboratory conditions.
Main Results:
- Metal-on-metal TMJ implants demonstrated volumetric wear an order of magnitude lower than acrylic-on-metal implants in laboratory tests.
- The wear rate of metal-on-metal implants was approximately half that reported for polyethylene-on-metal implants.
- Retrieved metal-on-metal implants showed abrasive wear with smaller wear zones compared to acrylic-on-metal implants.
Conclusions:
- Metal-on-metal bearing surfaces offer superior wear resistance for temporomandibular joint implants.
- Further research into wear characterization and minimization is essential for advancing TMJ arthroplasty.
- These findings support the continued development of metal-on-metal designs for improved TMJ implant longevity.