RF heating due to conductive wires during MRI depends on the phase distribution of the transmit field
Christopher J Yeung1, Robert C Susil, Ergin Atalar
1Department of Biomedical Engineering, Johns Hopkins University School of Medicine, Baltimore, Maryland 21205, USA.
Abstract:
In many studies concerning wire heating during MR imaging, a "resonant wire length" that maximizes RF heating is determined. This may lead to the nonintuitive conclusion that adding more wire, so as to avoid this resonant length, will actually improve heating safety. Through a theoretical analysis using the method of moments, we show that this behavior depends on the phase distribution of the RF transmit field. If the RF transmit field has linear phase, with slope equal to the real part of the wavenumber in the tissue, long wires always heat more than short wires. In order to characterize the intrinsic safety of a device without reference to a specific body coil design, this maximum-tip heating phase distribution must be considered. Finally, adjusting the phase distribution of the electric field generated by an RF transmit coil may lead to an "implant-friendly" coil design.
Related Concept Videos
Magnetic Field Due to Two Straight Wires
Magnetic Resonance Imaging
Magnetic Field Due To A Thin Straight Wire
Magnetic Force On Current-Carrying Wires: Example
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Inductance: Single-Phase And Three-Phase Line
Single-Phase Two-Wire Line:
A single-phase line consists of two solid cylindrical conductors, denoted as x and y. Each conductor carries phasor currents ix and iy, respectively. Given that the sum of these currents is...