Related Experiment Video
Updated: Sep 28, 2026

Frequency Mixing Magnetic Detection Scanner for Imaging Magnetic Particles in Planar Samples
Published on: June 9, 2016
Simulation of a birdcage and a ceramic cavity HF-resonator for high magnetic fields in magnetic resonance imaging
E Eriksen1, M A Golombeck, S Junge
1Institut für Biomedizinische Technik, IBT, Universität Karlsruhe (TH), Deutschland. eriksen@web.de, mag@ibt.uka.de
Abstract:
The aim of this work was the 3D-simulation of a dielectric resonator for high-field-MRI. A 12-rod-bird-cage-resonator was simulated in a first step, in order to verify the capability of the commercial simulation software MAFIA to simulate homogeneous, transversal B-fields in resonators. The second step was the simulation of frequency-independent dielectric ceramic resonators for static magnetic field strengths of 7 T and 12 T (294 MHz and 504 MHz respectively). The results were compared to the measured results of a manufactured TiO2- and a Al2O3-resonator. Only minor deviations showed up. These results led to the conclusion that dielectric resonators for high field MRI can be optimised using numerical field calculation software.
Related Concept Videos
Magnetic Resonance Imaging
Atomic Nuclei: Magnetic Resonance
NMR Spectrometers: Resolution and Error Correction
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...