Application of pixel truncation to reduce intensity artifacts in myocardial SPECT imaging with Tc-99m tetrofosmin
Masao Funahashi1, Tsuyoshi Shimonagata, Kazuhiro Mihara
1Department of Diagnostic Imaging, Osaka Prefectural General Hospital, Japan.
Background:
Technetium 99m-labeled radiopharmaceuticals accumulate in the liver and gallbladder, where they generate intensity artifacts that can result in misdiagnosis of myocardial single photon emission computed tomography (SPECT) images. This study identifies and eliminates factors affecting the magnitude and appearance of intensity artifacts in a gallbladder-heart phantom.
Methods And Results:
The myocardium and background compartments of a phantom were filled with Tc-99m at concentrations of 320 and 26.1 kBq/mL, respectively. A disposable plastic syringe containing 5 mL of Tc-99m as a model of the gallbladder was fixed in a position lateral to the heart phantom. Artifact intensity was determined on SPECT images over a specific activity range in the syringe (28.6, 6.6, and 0.2 MBq/mL). Among 72 projection images, those with maximal heart counts in the region of interest were selected. Counts above and below 110% of the maximal heart count in all projection images were excluded and reconstructed, respectively. At 28.6 and 6.6 MBq/mL, excessive artifacts generated cold pixels immediately around the source, whereas lower activity (0.2 MBq/mL) caused the artifacts to disappear. Truncating the counts in the gallbladder caused the intensity artifacts at specific activities of 28.6 and 6.6 MBq/mL to disappear.
Conclusions:
The magnitude and appearance of intensity artifacts depend on contrast between extracardiac activities in the same slices of the heart in myocardial SPECT images with Tc-99m tetrofosmin, and pixel truncation can eliminate them.
More Related Videos
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
11:17A Microscopic 2,3,5-Triphenyltetrazolium Chloride Assay for Accurate and Reliable Analysis of Myocardial Injury
Published on: November 28, 2025
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT