Chemical dosimetry using an iodide/iodate aqueous solution: application to the gamma irradiation of blood

Ronald O Rahn1

  • 1Department of Environmental Health, University of Alabama at Birmingham, Birmingham, AL 35294, USA. rahn@ms.soph.uab.edu

A method is presented for measuring and verifying the radiation dose in gamma irradiators used for treating blood prior to transfusion. This method employs the iodide/iodate dosimeter (0.6M iodide, 0.1M iodate, and 0.01 borate at pH 9.25) which forms triiodide upon exposure to ionizing radiation; for Cs-137 radiation the G value is 14.1. Samples were placed in a canister and irradiated in a conventional blood bank irradiator containing several Cs-137 sources. The following were exposed: (a) nine 1.5 ml plastic tubes containing dosimetry solution taped inside a 250 ml blood bag, which, in turn, was filled with either water or blood, (b) 50 ml plastic syringes containing varying amounts of dosimetry solution, (c) a whole blood bag containing 250 ml of the dosimetry solution. A water phantom was not used during exposure. The absorbance changes at 352 nm due to triiodide formation were used to determine a dose rate, which was on the order of 10 Gy/min (+/-5%) for all samples measured. This value is consistent with an average time-decayed dose rate for the irradiation volume as determined from the manufacturers calibration of the unit taking into account the heterogeneous nature of the radiation field inside the irradiator and the absence of a water phantom. Because of its sensitivity, ease of operation, and reproducibility, it is suggested that the iodide/iodate dosimetry system be considered for on-site periodic conformation/verification of the radiation dose as part of a quality assurance requirement for blood irradiators.

Related Concept Videos

Biological Effects of Radiation02:59

Biological Effects of Radiation

All radioactive nuclides emit high-energy particles or electromagnetic waves. When this radiation encounters living cells, it can cause heating, break chemical bonds, or ionize molecules. The most serious biological damage results when these radioactive emissions fragment or ionize molecules. For example, α and β particles emitted from nuclear decay reactions possess much higher energies than ordinary chemical bond energies. When these particles strike and penetrate matter, they produce ions...
Redox Titration: Overview01:21

Redox Titration: Overview

Redox titration is a chemical analysis technique used to determine the concentration of an unknown substance by measuring the electron transfer in a redox (reduction-oxidation) reaction. The process involves gradually adding a titrant with a known concentration of an oxidizing or reducing agent, to the analyte, the solution with an unknown concentration, until reaching the endpoint, which indicates the completion of the reaction between the two substances. Ensuring the analyte is in a single...
Redox Titration: Iodimetry and Iodometry01:23

Redox Titration: Iodimetry and Iodometry

Iodometry and iodimetry are analytical methods used to determine the concentration of oxidizing or reducing agents using iodine. In iodometric titrations, the oxidizing analyte solution is usually acidified and treated with an excess of iodide ions, which generates an equivalent amount of iodine in equilibrium with triiodide. The released iodine is subsequently titrated directly against a standardized reducing agent. As the dilute iodine color becomes pale yellow, a few drops of freshly...
EDTA: Indirect and Alkalimetric Titration01:23

EDTA: Indirect and Alkalimetric Titration

Unlike direct titration, back-titration, and displacement titration, indirect titration is an EDTA titration method for quantifying anions. In the indirect titration method, anions are precipitated as their insoluble salts with excess metal ions. The filtrate containing the excess metal ions is directly titrated with standard EDTA until the endpoint is achieved. Another approach involves extracting the metal ion and back-titrating with standard EDTA to obtain the endpoint. In this way, the...
Controlled-Current Coulometry: Overview01:27

Controlled-Current Coulometry: Overview

Controlled current coulometry, also known as amperostatic coulometry, is a technique used in electrochemical analysis to measure the quantity of a substance through the controlled passage of current. It involves the application of a constant current to an electrochemical cell containing the analyte of interest. As the current flows through the cell, the analyte undergoes a redox reaction at the electrode surface, resulting in a charge transfer. By monitoring the time required for a certain...
Imaging Studies II: Positron Emission Tomography and Scintigraphy01:25

Imaging Studies II: Positron Emission Tomography and Scintigraphy

Positron Emission Tomography (PET) is a medical imaging technique that provides crucial insights into the body's physiological functions at a molecular level. It is an indispensable resource for diagnosing, staging, and monitoring various illnesses, notably cancer, neurological disorders, and cardiovascular conditions.
Fundamental Principles of PET