Magnetic Adjustment of Afterload in Engineered Heart Tissues
Published on: May 5, 2020
Experimental method to reveal the effect of rotor magnet size and air gap on artificial heart driving motor torque
1Institute of Biomedical Engineering, University of Jiangsu at Zhenjiang, 212013, China.
Abstract:
To investigate experimentally the effect of rotor magnet design on artificial heart driving motor performance, seven rotors with different magnet lengths or thicknesses, as well as different peripheral angles, were manufactured and tested in the same motor stator with different rotating speeds. The input power (voltage and current) and output torque were measured and the motor efficiency was computed. The results demonstrated that the reduction of rotor magnet size and the enlargement of the air gap between the rotor magnets and the stator coil core have no significant effect on motor efficiency, but will reduce the torque value on which the motor achieves the highest efficiency; it could be remedied however by increasing the rotating speed, because the torque at the high efficiency point will increase along with the rotating speed. These results may provide a basis for developing small rotor magnets, large air gap and high efficiency motors for driving an artificial heart pump.
Related Concept Videos
Electro-mechanical Systems
A key component of the DC motor is the armature, a rotating circuit positioned within a magnetic field. As an electric current passes through the...
Force On A Current Loop In A Magnetic Field
Torque On A Current Loop In A Magnetic Field
Consider a rectangular current-carrying loop containing N turns of wire, placed in a uniform magnetic field. The net force on a current-carrying loop...
Back EMF
Electric Generator: Alternator
The magnetic flux passing through the coil varies sinusoidally as the loop rotates inside the magnetic field. This...
Faraday Disk Dynamo