Vascular Balloon Injury and Intraluminal Administration in Rat Carotid Artery
Published on: December 23, 2014
Microembolization during carotid stenting with the distal-balloon antiemboli system
N Al-Mubarak1, G S Roubin, J J Vitek
1Endovascular Therapeutics, University Hospital System of Cleveland, OH 44145, USA. NALmubarak@aol.com
Background:
The distal-balloon protection system is being evaluated for its efficacy in preventing embolic neurological events during carotid stenting (CAS). We sought to determine the effect of this system on the frequency of the Doppler-detected microembolic signals (MES) during CAS.
Methods:
Using transcranial Doppler, we compared the frequency of MES during CAS in 2 groups; 39 patients without distal protection and 37 with the distal-balloon protection system (GuardWire, Percusurge, Sunnyville, CA). There was no significant difference in the clinical or angiographic characteristics between the 2 groups. Three phases with increased MES counts were identified during the unprotected CAS; stent deployment, predilation, and postdilation (mean+/-SD: 75+/-57, 32+/-36, and 27+/-25, respectively).
Results:
The distal-balloon protection significantly reduced the frequency of MES during CAS (MES-counts: 164+/-108 in the control vs 68+/-83 in the protection group, p=0.002) particularly during these 3 phases. MES in the protection group were detected predominantly during sheath placement, guidewire manipulation and during the distal-balloon deflation.
Conclusions:
Three phases with increased MES counts were identified during the unprotected CAS, e.g. stent deployment, predilation and postdilation. The distal-balloon protection system significantly reduced the frequency of MES during CAS, particularly during these 3 phases.