Cardiac positioning using an apical suction device maintains beating heart hemodynamics
Jerome Sepic1, Jon O Wee, Edward G Soltesz
1Division of Cardiac Surgery, Brigham and Women's Hospital, Harvard Medical School, Boston, MA, USA. jsepic@partners.org
The Heart Surgery Forum
|January 23, 2003
Summary
Deep pericardial sutures (DPS) for cardiac positioning during off-pump coronary artery bypass (OPCAB) cause significant hemodynamic compromise. An apical suction device offers superior cardiac positioning with minimal hemodynamic impact.
Area of Science:
- Cardiovascular Surgery
- Surgical Devices
- Hemodynamics
Background:
- Cardiac positioning is crucial in off-pump coronary artery bypass (OPCAB).
- Deep pericardial sutures (DPS) are commonly used but can lead to hemodynamic compromise.
- Alternative methods for cardiac positioning are needed to improve patient outcomes.
Purpose of the Study:
- To compare the hemodynamic effects of an apical suction device versus deep pericardial sutures (DPS) for cardiac positioning during OPCAB.
- To evaluate the efficacy of an apical suction device in maintaining hemodynamic stability.
Main Methods:
- A study was conducted on five healthy pigs undergoing sternotomy.
- Hemodynamic parameters including right atrial pressure, left ventricular pressure and volume, and aortic pressure and flow were measured.
- Measurements were taken at baseline, with the apical suction device, and during cardiac positioning using both DPS and the apical suction device.
Main Results:
- Deep pericardial sutures (DPS) significantly compromised hemodynamics, reducing cardiac output by up to 30% and mean arterial pressure by up to 26%.
- DPS also led to decreased left ventricular end-diastolic pressure and increased right atrial pressure.
- The apical suction device demonstrated minimal hemodynamic compromise, with only a modest decrease in cardiac output and right atrial pressure during exposure.
Conclusions:
- Deep pericardial sutures (DPS) significantly impair hemodynamics due to reduced left ventricular filling.
- The apical suction device provides adequate cardiac exposure with significantly less hemodynamic compromise compared to DPS.
- An apical suction device represents a hemodynamically superior alternative for cardiac positioning in OPCAB procedures.