Related Experiment Video
Updated: Jul 30, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Comparison of computed radiography and film/screen combination using a contrast-detail phantom
Z F Lu1, E L Nickoloff, J C So
1Department of Radiology, Columbia University, MHB 3-265, 177 Ft. Washington Avenue, New York, New York 10032, USA. zfl1@columbia.edu
This study compared computed radiography (CR) and film/screen systems, finding that higher kVp and added aluminum filtration reduce patient dose without harming image quality. Soft-copy display aids contrast-detail detection.
Area of Science:
- Medical Imaging Physics
- Radiologic Technology
Background:
- Digital imaging technologies like computed radiography (CR) aim to improve diagnostic accuracy and patient safety.
- Traditional film/screen radiography remains a benchmark for image quality assessment.
Purpose of the Study:
- To compare Kodak CR 400 and Speed 400 film/screen systems.
- Evaluate patient dose, technique settings, and contrast-detail detectability.
- Assess the impact of added filtration on CR performance.
Main Methods:
- Utilized a contrast-detail phantom with varying hole diameters and depths.
- Acquired images using film/screen and CR systems across a range of kVp (60-120 kVp).
- Measured patient entrance skin dose and evaluated image quality via detection ratios.
Main Results:
- Adding 2 mm aluminum filtration to the CR system reduced patient entrance skin dose.
- Higher kVp settings also contributed to decreased patient dose.
- Contrast-detail detectability was maintained, especially with soft-copy display adjustments.
Conclusions:
- Optimizing technique factors (higher kVp) and using added filtration can lower patient dose in CR imaging.
- Digital radiography offers flexibility in image display, compensating for potential contrast changes.
- CR systems can achieve comparable or improved diagnostic performance over film/screen with appropriate settings.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography