Related Experiment Video
Updated: Aug 19, 2026

Online Size-exclusion and Ion-exchange Chromatography on a SAXS Beamline
Published on: January 5, 2017
[Determination of ibandronate by high performance ion exchange chromatography]
1National Laboratory of Nuclear Medicine, Jiangsu Institute of Nuclear Medicine, Wuxi 214063, China.
Abstract:
A method for the determination of Ibandronate by high performance ion exchange chromatography is reported. Ibandronate can form a complex compound with the copper ion added in the mobile phase. The complex can be detected at UV 240 nm. The effects of pH of mobile phase and concentration of methanol to the retention time were studied. Concentration of 10% methanol and pH 2.5 were suitable for the detection in less than 5 min. The flow rate was 1.0 mL/min. The relative standard deviation of peak area was 0.8%. The linear correlation was observed from 300 mg/L to 1,500 mg/L of Ibandronate (r = 0.9989). The lowest detectable concentration of Ibandronate was 6 mg/L. The method is simple, rapid and reproducible.
More Related Videos
07:54Analysis of AtHIRD11 Intrinsic Disorder and Binding Towards Metal Ions by Capillary Gel Electrophoresis and Affinity Capillary Electrophoresis
Published on: August 22, 2018
10:41Ion Exchange Chromatography (IEX) Coupled to Multi-angle Light Scattering (MALS) for Protein Separation and Characterization
Published on: April 5, 2019
Related Concept Videos
Types Of Column Chromatography
Gel Filtration Chromatography
When the...
Ion Exchange
Ion-Exchange Chromatography