[Determination of blood glucose by chemiluminescent flow-injection analysis with immobilized glucoxidase column]
1Research Lab of Applied Chemistry, School of Pharmacy, WCUMS, Chengdu 610041.
Abstract:
Luminol-hemin system was used for chemiluminescent flow-injection analysis with immobilized glucoxidase(GOD) column to determine blood glucose. The method is fast, simple and sensitive. Blood samples were diluted to 1:1000 and pumped into the flow-injection analyzer for direct detection. The sampling frequency was 30 samples/h, with a detection limit of 7.2 x 10(-9) mol/L(1.30 micrograms/L). A column packed with 36 milligrams of CPG could be used nearly 200 times. The determination results of glucose in serum by this method were in good agreement with those obtained by the routine GOD spectrophotometric method (correlation coefficient r = 0.90, P > 0.05). Recoveries of this method were 94.0%-107.2%.