Related Experiment Video
Updated: Aug 11, 2026

Phase-Resolved Functional Lung MRI for Pulmonary Ventilation and Perfusion (V/Q) Assessment
Published on: August 9, 2024
Estimation of regional lung volume changes by electrical impedance pressures tomography during a pressure-volume
Huibert R van Genderingen1, Adrianus J van Vught, Jos R C Jansen
1Department of Clinical Physics and Informatics, Vrije Universiteit Medical Center, PO Box 7057, 1007 MB, Amsterdam, The Netherlands. hr.genderingen@vumc.nl
Objective:
To assess the degree of linearity between lung volume and impedance change by electrical impedance tomography (EIT) in pigs with acute lung injury and to investigate regional impedance changes during a pressure-volume maneuver.
Design And Setting:
Experimental animal study in a university research laboratory.
Patients And Participants:
Nine pigs with lung injury induced by lung lavage.
Interventions:
The lungs were insufflated to four different lung volumes. Next the lungs were inflated in steps up to 40 cm H(2)O and then in steps deflated.
Measurements And Results:
EIT measurements were performed. Impedance was highly linear with lung volume ( r(2)=0.97). From the pressure-volume maneuver regional pressure-impedance (P-I) curves were obtained in the upper half (ventral) and lower half (dorsal) of the thoracic cross-section. Excellent fit was found of the regional P-I curves with a predefined sigmoid equation ( r(2)=0.998). The P-I curves after lavage were markedly different than before lavage. The P-I curves recorded after lavage displayed a strong heterogeneity on the inflation limb: Lower corner pressure (traditionally lower inflection point) was significantly higher in the dorsal (28.3+/-4.1 cm H(2)O) than in the ventral region (17.5+/-4.3 cm H(2)O). The deflation limb displayed a more homogeneous pattern. Upper corner pressure and true inflection point, where the curve slope is maximal, in the dorsal region were only slightly higher than in the ventral region (1-2 cm H(2)O).
Conclusions:
EIT and automated curve fitting provide information on regional lung inflation and deflation which may be of clinical use for optimizing ventilator settings.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...