Evaluation of Drug Sorption to PVC- and Non-PVC-based Tubes in Administration Sets Using a Pump
Published on: March 11, 2017
Lost bits: particle shedding with polyvinyl chloride intravenous administration sets
Peter Subramanian1, Paul Ainsworth, John Cassey
1Department of Cardiothoracic surgery, John Hunter Hospital, Newcastle, NSW, Australia.
Abstract:
Silicone particles have been demonstrated in the effluent from silicone intravenous (IV) tubing. It has been widely suspected that polyvinyl chloride (PVC) particles are also lost. We sought to clarify the situation in a carefully controlled laboratory setting using the apparatus and flow rates common in a paediatric setting, using scanning electron microscope techniques (SEM), we found that particles were indeed shed from IV tubing during use, but they were not PVC.