Related Experiment Video
Updated: Aug 12, 2026

Enhanced Genetic Analysis of Single Human Bioparticles Recovered by Simplified Micromanipulation from Forensic ‘Touch DNA’ Evidence
Published on: March 9, 2015
[Effect of PCR reaction volume on the accuracy of human identification tests]
Huai-gu Zhou1, Yuan Ping, Qing-wen Xu
1Shanghai Institute of Criminal Science and Technology, Shanghai 200083. hgzhou@public2.sta.net.cn
Objective:
To investigate the effect of different PCR amplification volume on the accuracy of human identification test.
Methods:
Human genome DNA samples were amplified using ABI PRISM Profiler Plus kits in 50 microliters, 25 microliters, 12.5 microliters, and 6.25 microliters reaction volume, respectively. The thermocycle parameters were the same. All PCR products were then electrophoresized on ABI PRISM 310 Genetic Analyzer, 377 DNA Sequencer, and 3100 Genetic Analyzer. Data were processed by ABI PRISM GeneScan and Genotyper software.
Results:
The less reaction volume, the more alleles losing or alleles adding observed.
Conclusion:
Non-standard volume of PCR amplification reaction should be used carefully in human identification test, especially when the sample DNA quality is not so satisfied.
More Related Videos
05:58Digital Polymerase Chain Reaction Assay for the Genetic Variation in a Sporadic Familial Adenomatous Polyposis Patient Using the Chip-in-a-tube Format
Published on: August 20, 2018
08:54Development and Testing of Species-specific Quantitative PCR Assays for Environmental DNA Applications
Published on: November 5, 2020
Related Concept Videos
PCR
PCR - Polymerase Chain Reaction
Real Time RT-PCR
The real-time quantification of the number of amplified products is...