Influence of phantom diameter, kVp and scan mode upon computed tomography dose index
Related Experiment Video
Updated: Sep 26, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Edward L Nickoloff1, Ajoy K Dutta, Zheng F Lu
1Department of Radiology, Columbia University, New York, New York 10962, USA. eln1@columbia.edu
Insights
Computed tomography (CT) radiation dose varies significantly with phantom size and kVp. This study develops equations to estimate CT dose index (CTDI) for diverse patient sizes and scan conditions, improving pediatric radiation dose assessments.
Area of Science:
- Medical Physics
- Radiological Sciences
- Imaging Technology
Background:
- Pediatric computed tomography (CT) radiation dose is a growing concern.
- Accurate CT dose index (CTDI) estimation across various patient sizes is crucial.
- Current protocols use limited phantom sizes, necessitating broader applicability.
Purpose of the Study:
- To investigate the relationship between CTDI, phantom size, kVp, and scan mode.
- To develop predictive equations for CTDI based on these parameters.
- To provide a method for estimating radiation doses for diverse CT scenarios.
Main Methods:
- Measurements of CTDI were performed using phantoms ranging from 6 cm to 32 cm in diameter.
- X-ray tube potentials varied from 80 to 140 kVp.
- Consecutive axial, single-slice helical, and multislice helical scan modes were employed.
Main Results:
- CTDI demonstrates an exponential dependence on phantom diameter.
- Smaller phantoms receive significantly higher radiation doses than larger ones for equivalent kVp and mAs.
- Simplified equations were derived to relate phantom diameter and kVp to CTDI.
Conclusions:
- The established relationships allow for the estimation of CTDI across a wide range of patient sizes and scan parameters.
- Findings can inform more accurate radiation dose assessments, particularly for pediatric patients.
- A method for converting axial CT dose measurements to helical scan values was also presented.
Abstract:
The computed tomography (CT) radiation dose to pediatric patients has received considerable attention recently. Moreover, it is important to be able to determine CT radiation doses for various patient sizes ranging from infants to large adults. The current AAPM protocol only measures CT radiation dose using a 16 cm acrylic phantom to represent an adult head and a 32 cm acrylic phantom to represent an adult body. The goal of this paper is to study the dependence of the computed tomography dose index (CTDI) upon the size of the phantom, the kVp selected and the scan mode employed. Our measurements were done on phantom sizes ranging from 6 cm to 32 cm. The x-ray tube potential ranged from 80 to 140 kVp. The scan modes utilized for the measurements included: consecutive axial scans, single-slice helical scans with variable pitch and multislice helical scans with variable pitch. The results were consolidated into simplified equations which related the phantom diameter and kVp to the measured CTDI. Some generalizations were made about the relationship between the scan modes of the various CT units to the measured radiation doses. The CTDI appears to be an exponential function of phantom diameter. For the same kVp and mAs, the radiation doses for smaller phantoms are much greater than for larger sizes. The derived relationship can be used to estimate the radiation doses for a variety of scan conditions and modes from measurements with the two standard reference phantoms. A method was also given for converting axial CT dose measurements to appropriate MSAD values for helical CT scans.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT