Comparison of continuous cardiac output measurements in patients after cardiac surgery
Frank Mielck1, Wolfgang Buhre, Gunnar Hanekop
1Departments of Anesthesiology, and Medical Statistics, University of Goettingen, Germany. fmielck@gwdg.de
Objective:
To investigate in a direct comparison accuracy and precision of continuous cardiac output measurements assessed by continuous pulmonary artery thermodilution technique (TDCCO), continuous pulse contour analysis (PCCO), and noninvasive partial CO(2)-rebreathing technique (NICO) in patients after coronary artery bypass grafting (CABG) during the postoperative period.
Design:
Prospective, controlled clinical study.
Setting:
University hospital.
Participants:
Twenty-two patients undergoing elective CABG surgery.
Interventions:
Hemodynamic measurements were performed after admission to the ICU and in sequence every 2 hours during the subsequent 6-hour period. Simultaneously, cardiac output (CO) was measured using a TDCCO, PCCO, and NICO. After the continuous cardiac output measurements were read, bolus thermodilution-derived cardiac output was obtained from thermodilution curves detected in the pulmonary artery (TDBCO(pa)). Four intermittent consecutive boli consisting of 10 mL of ice-cold saline were randomly injected over the ventilatory cycle.
Measurements And Main Results:
The comparison between the continuous cardiac output measurement methods TDCCO versus PCCO showed a bias of -0.12 L/min, between TDCCO versus NICO -0.17 L/min, and between PCCO versus NICO -0.44 L/min. The comparison to the reference technique between TDBCO(pa) versus TDCCO revealed a bias of -0.28 L/min, between TDBCO(pa) versus PCCO -0.40 L/min, and between TDBCO(pa) versus NICO -0.64 L/min.
Conclusions:
The results of this clinical investigation show agreement between TDCCO and PCCO to satisfy clinical requirements in a setting of postoperative patients after cardiac surgery. In contrast, the NICO monitor is of very limited use in these patients.
Related Concept Videos
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiomyopathy VII: Pre and Post Operative Nursing Management