Morin was used as a pre-column reagent for the determination of aluminum by RP-HPLC with fluorescence detection. This method has been successfully applied to direct determination of trace Al in environmental and biological samples. The response was linear from 6 x 10(-9) to 6 x 10(-5) M with a detection limit of 2 x 10(-9) M. In addition, the different Al complexes with morin were separated by the proposed HPLC procedure and their coordination ratios were depicted by molar-ratio method. The results showed that 1:1 and 2:1 Al-morin complexes formed.