Reducing vascular access morbidity: a comparative trial of two vascular access monitoring strategies

Charmaine E Lok1, Cynthia Bhola, Ruth Croxford

  • 1Department of Medicine, University Health Network, The Toronto General Hospital, Toronto, Canada. charmaine.lok@uhn.on.ca

Insights

The saline ultrasound dilution technique (Transonic) significantly reduced access thrombosis in dialysis patients compared to Duplex ultrasonography. This new monitoring strategy lowered thrombosis rates and decreased the need for procedures like angioplasty.

Area of Science:

  • Vascular access monitoring in hemodialysis
  • Clinical effectiveness of monitoring techniques
  • Thrombosis prevention strategies

Background:

  • Thrombosis is a major cause of failure for hemodialysis grafts and fistulas, leading to significant patient morbidity.
  • Current interventions focus on maintaining access patency through early detection and treatment of lesions.
  • The optimal monitoring strategy for preventing access thrombosis remains unclear.

Purpose of the Study:

  • To compare the efficacy of Duplex ultrasonography and the saline ultrasound dilution technique (Transonic) in preventing access thrombosis.
  • To identify risk factors associated with access thrombosis.
  • To evaluate the performance of two distinct monitoring strategies over a three-year period.

Main Methods:

  • A prospective, three-year study in a large dialysis center comparing monitoring strategies.
  • Year 1: Duplex ultrasonography; Year 3: Transonic monitoring (Year 2: transition).
  • Multivariate survival analysis was used to determine thrombosis risk factors and assess monitoring technique performance.

Main Results:

  • The total thrombosis rate decreased from 1.01/1000 access days in Year 1 to 0.66/1000 access days in Year 3.
  • This reduction was achieved with decreased rates of angiograms (55%), angioplasties (13%), and thrombolysis (31%).
  • Low flow rates detected by Transonic monitoring correlated with increased thrombosis risk.

Conclusions:

  • The saline ultrasound dilution technique (Transonic) demonstrated superior effectiveness in reducing access thrombosis compared to Duplex ultrasonography.
  • Stenosis identified by Duplex ultrasonography was a weaker predictor of impending thrombosis.
  • Different monitoring techniques may be best suited for specific clinical scenarios in managing vascular access.