A comparison of four projection perimeters
American Journal of Ophthalmology
|May 1, 1976
Summary
Different perimeters showed varied visual field mapping capabilities. The Goldmann perimeter was most effective, while the Rodenstock struggled, likely due to variations in target size and luminance.
Area of Science:
- Ophthalmology
- Visual Science
- Medical Devices
Background:
- Perimetry is crucial for diagnosing visual field defects.
- Standardized specifications for perimeters suggest comparable performance.
- Variations in instrument performance can impact clinical diagnosis.
Purpose of the Study:
- To compare the performance of four projection perimeters: Goldmann, Marco, Topcon, and Rodenstock.
- To identify differences in visual field mapping capabilities among these instruments.
- To investigate the reasons behind observed performance variations.
Main Methods:
- Thirty eyes were subjected to visual field testing using all four perimeters.
- Standardized targets and dome specifications were used initially.
- Subsequent analysis examined actual target size and luminance for each machine.
Main Results:
- The Rodenstock perimeter frequently failed to map central fields or detect defects.
- The Goldmann perimeter demonstrated the highest sensitivity and definition in mapping scotomata.
- Topcon and Marco perimeters showed intermediate sensitivity, with Topcon slightly outperforming Marco.
Conclusions:
- Despite identical specifications, significant variations in target size and luminance exist among perimeters.
- These variations directly impact the ability to accurately plot visual fields and detect defects.
- The Goldmann perimeter offers superior performance for visual field assessment compared to the tested alternatives.