Assessment of Neuromuscular Function Using Percutaneous Electrical Nerve Stimulation
Published on: September 13, 2015
Simple anatomical measurements do not correlate significantly to individual peripheral nerve stimulation thresholds
Blaine A Chronik1, Meena Ramachandran
1Department of Electrical Engineering, Stanford University, Stanford, California 94305, USA. bchronik@stanford.edu
Purpose:
To examine peripheral nerve stimulation (PNS) thresholds for normal human subjects in magnetic resonance imaging (MRI) gradient coils, and determine if observed thresholds could be predicted based on gross physiologic measurements.
Materials And Methods:
PNS thresholds for 21 healthy normal subjects were measured using a whole-body gradient coil. Subjects were exposed to a trapezoidal echo-planar imaging (EPI) gradient waveform and the total change in gradient strength (DeltaG) required to cause PNS as a function of the duration of the gradient switching time (tau) were measured. Correlation coefficients and corresponding P values were calculated for the PNS threshold measurements against simple physiologic measurements taken of the subjects, including weight, height, girth, and average body fat percentage, in order to determine if there were any easily observable dependencies.
Results:
No convincing correlations between threshold parameters and gross physiologic measurements were observed.
Conclusion:
These results suggest it is unlikely that a simple physiologic measurement of subject anatomy can be used to guide the operation of MRI scanners in a subject-specific manner in order to increase gradient system performance while avoiding PNS.