TG-51: experience from 150 institutions, common errors, and helpful hints
R C Tailor1, W F Hanson, G S Ibbott
1Department of Radiation Physics, University of Texas M.D. Anderson Cancer Center, 1515 Holcombe Blvd., Box 547, Houston, Texas 77030, USA.
Abstract:
The Radiological Physics Center (RPC) is a resource to the medical physics community for assistance regarding dosimetry procedures. Since the publication of the AAPM TG-51 calibration protocol, the RPC has responded to numerous phone calls raising questions and describing areas in the protocol where physicists have had problems. At the beginning of the year 2000, the RPC requested that institutions participating in national clinical trials provide the change in measured beam output resulting from the conversion from the TG-21 protocol to TG-51. So far, the RPC has received the requested data from approximately 150 of the approximately 1300 institutions in the RPC program. The RPC also undertook a comparison of TG-21 and TG-51 and determined the expected change in beam calibration for ion chambers in common use, and for the range of photon and electron beam energies used clinically. Analysis of these data revealed two significant outcomes: (i) a large number (approximately 1/2) of the reported calibration changes for photon and electron beams were outside the RPC's expected values, and (ii) the discrepancies in the reported versus the expected dose changes were as large as 8%. Numerous factors were determined to have contributed to these deviations. The most significant factors involved the use of plane-parallel chambers, the mixing of phantom materials and chambers between the two protocols, and the inconsistent use of depth-dose factors for transfer of dose from the measurement depth to the depth of dose maximum. In response to these observations, the RPC has identified a number of circumstances in which physicists might have difficulty with the protocol, including concerns related to electron calibration at low energies (R50<2 cm), and the use of a cylindrical chamber at 6 MeV electrons. In addition, helpful quantitative hints are presented, including the effect of the prescribed lead filter for photon energy measurements, the impact of shifting the chamber depth for photon depth-dose measurements, and the impact of updated stopping-power data used in TG-51 versus that used in TG-21, particularly for electron calibrations.
More Related Videos
Related Concept Videos
Types of Errors: Detection and Minimization
Absolute error in a measurement is the numerical difference from the true or central value. Relative error is the ratio between absolute error and the true or central value, expressed as a percentage.
Errors can be classified by source, magnitude, and sign. There are three types of errors: systematic, random, and gross.
Systematic or...
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5% chance...
Errors and Mistakes in Surveying
Common Leveling Mistakes and Errors