Related Experiment Videos
Three-dimensional visual guidance improves the accuracy of calculating right ventricular volume with two-dimensional
Jennifer L Dorosz1, Edward L Bolson, Mary S Waiss
1Cardiovascular Research and Training Center, University of Washington, Box 356422, Seattle, WA 98195, USA. dorosz@u.washington.edu
Abstract:
Three-dimensional guidance programs have been shown to increase the reproducibility of 2-dimensional (2D) left ventricular volume calculations, but these systems have not been tested in 2D measurements of the right ventricle. Using magnetic fields to identify the probe location, we developed a new 3-dimensional guidance system that displays the line of intersection, the plane of intersection, and the numeric angle of intersection between the current image plane and previously saved scout views. When used by both an experienced and an inexperienced sonographer, this guidance system increases the accuracy of the 2D right ventricular volume measurements using a monoplane pyramidal model. Furthermore, a reconstruction of the right ventricle, with a computed volume similar to the calculated 2D volume, can be displayed quickly by tracing a few anatomic structures on 2D scans.