Analysis of Factors Influencing Accuracy of Volume Flow Measurement in Dialysis Access Fistulas Based on Duplex
R Eugene Zierler1, Daniel F Leotta2, Kurt Sansom3
11 Division of Vascular Surgery, Department of Surgery, University of Washington, Seattle, WA, USA.
Objective:
We developed a duplex ultrasound simulator and used it to assess accuracy of volume flow measurements in dialysis access fistula (DAF) models.
Methods:
The simulator consists of a mannequin, computer, and mock transducer. Each case is built from a patient's B-mode images that are used to create a 3-dimensional surface model of the DAF. Computational fluid dynamics is used to determine blood flow velocities based on model vessel geometry. The simulator displays real-time B-mode and color-flow images, and Doppler spectral waveforms are generated according to user-defined settings. Accuracy was assessed by scanning each case and measuring volume flow in the inflow artery and outflow vein for comparison with true volume flow values.
Results:
Four examiners made 96 volume flow measurements on four DAF models. Measured volume flow deviated from the true value by 35 ± 36%. Mean absolute deviation from true volume flow was lower for arteries than veins (22 ± 19%, N = 48 vs. 58 ± 33%, N = 48, p < 0.0001). This finding is attributed to eccentricity of outflow veins which resulted in underestimating true cross-sectional area. Regression analysis indicated that error in measuring cross-sectional area was a predictor of error in volume flow measurement (β = 0.948, p < 0.001). Volume flow error was reduced from 35 ± 36% to 9 ± 8% (p < 0.000001) by calculating vessel area as an ellipse.
Conclusions:
Duplex volume flow measurements are based on a circular vessel shape. DAF inflow arteries are circular, but outflow veins can be elliptical. Simulation-based analysis showed that error in measuring volume flow is mainly due to assumption of a circular vessel.
Related Concept Videos
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision
Bioavailability: Influencing Factors
Factors Influencing Microbial Growth: pH
Factors Influencing Attraction I: Proximity
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...