Related Experiment Video
Updated: Aug 7, 2026

Failure of Cleaning Verification in Pharmaceutical Industry Due to Uncleanliness of Stainless Steel Surface
Published on: August 11, 2017
Inadequacy of manual cleaning for reprocessing single-use, triple-lumen sphinctertomes: simulated-use testing
Michelle J Alfa1, Radita Nemes
1Department of Clinical Laboratory Science, Wayne State University, Detroit, Michigan, USA.
Background And Aims:
Despite widespread reuse of single-use sphinctertomes, publications regarding the adequacy of reprocessing are conflicting and the cautery wire channel is seldom evaluated. Our objective was to use thickened artificial test soil containing microorganisms to perform simulated-use tests combined with in-situ and destructive testing to evaluate cleaning efficacy and ethylene oxide sterilization of single-use triple lumen sphinctertomes.
Methods:
New triple-lumen sphinctertomes were soiled with thickened artificial test soil containing 6 log(10) per milliliter of Enterococcus faecalis and Bacillus stearothermophilus by inoculation through the distal end and dried for 1 hour, 24 hours, or 7 days before cleaning. The efficacy of manual cleaning was compared with that of automated cleaning with the Medisafe SI-Auto narrow-lumen cleaner. After cleaning, Bradford's reagent was injected into the channels as a direct method of detecting residual protein. Destructive testing was done to determine the levels of residual protein, carbohydrate, hemoglobin, endotoxin, and viable bacteria in the cleaned device. Destructive sterility testing of the devices also was performed after ethylene oxide sterilization.
Results:
Both in-situ and destructive testing demonstrated that manual cleaning and automated washers connected via the luer ports did not remove soil or organisms from the cautery wire channel. Only retro-flushing in the SI-Auto provided adequate cleaning of all 3 channels. If reprocessing was delayed for more than 24 hours, retro-flush cleaning was no longer effective. Ethylene oxide sterilization failure was detected only for devices held for 7 days before cleaning and sterilization. In-use testing showed that patient secretions gained access to the cautery wire channel.
Conclusions:
Only retro-flushing done within 24 hours of use provided adequate cleaning for multi-lumen, single-use sphinctertomes. Our data validate the efficacy of reprocessing sphinctertomes once with SI-Auto retro-flush cleaning followed by 2 hours of ethylene oxide sterilization.
Insights
Reprocessing single-use triple lumen sphinctertomes requires retro-flushing within 24 hours for effective cleaning. Ethylene oxide sterilization is validated when performed after successful retro-flush cleaning within this timeframe.
Area of Science:
- Gastroenterology
- Medical Device Reprocessing
- Infectious Disease Control
Background:
- Single-use triple lumen sphinctertomes are widely reused despite conflicting data on reprocessing effectiveness.
- The cautery wire channel's cleanliness is rarely assessed, posing a potential risk.
- Simulated-use testing is crucial to evaluate reprocessing protocols for complex medical devices.
Purpose of the Study:
- To assess the cleaning efficacy of reprocessing single-use triple lumen sphinctertomes using artificial test soil and microorganisms.
- To evaluate the effectiveness of manual versus automated cleaning methods, including retro-flushing.
- To determine the impact of reprocessing delay on cleaning and sterilization outcomes.
Main Methods:
- Triple-lumen sphinctertomes were soiled with artificial test soil and microorganisms, then dried for varying durations (1 hour to 7 days).
- Cleaning efficacy was evaluated using in-situ protein detection (Bradford's reagent) and destructive testing (protein, carbohydrate, hemoglobin, endotoxin, viable bacteria).
- Automated cleaning with Medisafe SI-Auto, including retro-flushing, was compared to manual cleaning; ethylene oxide sterilization was also assessed.
Main Results:
- Manual cleaning and standard automated washing failed to clean the cautery wire channel.
- Only retro-flushing with the SI-Auto effectively cleaned all three channels when performed within 24 hours of soiling.
- Reprocessing delayed beyond 24 hours rendered retro-flushing ineffective, and sterilization failed for devices soiled for 7 days.
Conclusions:
- Retro-flushing within 24 hours is essential for adequate reprocessing of multi-lumen, single-use sphinctertomes.
- Reprocessing with SI-Auto retro-flush cleaning followed by 2 hours of ethylene oxide sterilization is validated.
- Ensuring timely reprocessing is critical for patient safety when reusing single-use medical devices.
More Related Videos
10:29Semi-automatic PD-L1 Characterization and Enumeration of Circulating Tumor Cells from Non-small Cell Lung Cancer Patients by Immunofluorescence
Published on: August 14, 2019
03:10Pilot In Vitro Study to Assess Cleaning Ability and Effects of Different Decontamination Methods on Implant Surfaces
Published on: November 21, 2025
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Tracheostomy Suctioning II: Procedure
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...