Magnetic resonance imaging at 3.0 Tesla: challenges and advantages in clinical neurological imaging
Richard Frayne1, Bradley G Goodyear, Peter Dickhoff
1Seaman Family MR Research Centre, Foothills Medical Centre/University of Calgary, 1403 29th Street NW, Calgary, AB, Canada. rfrayne@ucalgary.ca
Abstract:
MR imaging at very high field (3.0 T) is a significant new clinical tool in the modern neuroradiological armamentarium. In this report, we summarize our 40-month experience in performing clinical neuroradiological examinations at 3.0 T and review the relevant technical issues. We report on these issues and, where appropriate, their solutions. Issues examined include: increased SNR, larger chemical shifts, additional problems associated with installation of these scanners, challenges in designing and obtaining appropriate clinical imaging coils, greater acoustic noise, increased power deposition, changes in relaxation rates and susceptibility effects, and issues surrounding the safety and compatibility of implanted devices. Some of the these technical factors are advantageous (eg, increased signal-to-noise ratio), some are detrimental (eg, installation, coil design and development, acoustic noise, power deposition, device compatibility, and safety), and a few have both benefits and disadvantages (eg, changes in relaxation, chemical shift, and susceptibility). Fortunately solutions have been developed or are currently under development, by us and by others, for nearly all of these challenges. A short series of 1.5 T and 3.0 T patient images are also presented to illustrate the potential diagnostic benefits of scanning at higher field strengths. In summary, by paying appropriate attention to the discussed technical issues, high-quality neuro-imaging of patients is possible at 3.0 T.
Related Concept Videos
Magnetic Resonance Imaging
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic Stimulation (TMS).
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan: