Microfracture of a baclofen pump catheter with intermittent under- and overdose
William J Dawes1, James M Drake, Darcy Fehlings
1Department of Neurosurgery, Hospital for Sick Children, and Bloorview-MacMillan Rehabilitation Center, Toronto, Canada.
Abstract:
An 8-year-old boy with an implanted intrathecal baclofen pump presented with intermittent signs of baclofen under- and overdose. Pump interrogation, pump volumes, plain X-rays, catheter contrast fluoroscopy and CT of the intrathecal catheter tip were unremarkable. A microfracture of the extradural catheter was only seen at a magnification of x20 and confirmed by scanning electron microscopy. Symptoms resolved following replacement. Microfracture of the intrathecal catheter of baclofen pump systems may present puzzling symptomatology and be difficult to diagnose by conventional means.
Insights
A microfracture in an intrathecal baclofen pump catheter caused confusing symptoms of under- and overdose. This rare catheter defect requires advanced imaging for diagnosis, impacting baclofen pump therapy.
Area of Science:
- Neurology
- Biomedical Engineering
- Medical Device Technology
Background:
- Intrathecal baclofen pump (IBP) therapy is crucial for managing spasticity.
- Catheter integrity is vital for effective and safe drug delivery.
- Complications can arise from implanted medical devices, necessitating thorough investigation.
Observation:
- An 8-year-old boy with an IBP experienced fluctuating symptoms suggestive of both baclofen under- and overdose.
- Standard diagnostic methods including pump interrogation, imaging (X-ray, CT), and fluoroscopy did not reveal the cause.
- A subtle extradural catheter microfracture was identified only upon high-magnification examination and scanning electron microscopy.
Findings:
- A previously undetected microfracture in the extradural portion of the intrathecal catheter was the source of the patient's symptoms.
- The microfracture led to unpredictable drug delivery, mimicking under- and overdose scenarios.
- Symptomatic resolution occurred after the catheter was surgically replaced.
Implications:
- Catheter microfractures in IBP systems can present diagnostic challenges.
- Advanced imaging techniques may be necessary to identify subtle catheter defects.
- This case highlights the importance of considering rare mechanical failures in IBP complications.