Threshold measurements: ten rules for good measuring practice.

Werner Irnich1

  • 1Department of Legal Medicine, University Hospital, Giessen, Germany. werner.irnich@technik.med.uni-giessen.de

Pacing and Clinical Electrophysiology : PACE
|July 25, 2003
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

A Model of Electrostimulation Based on the Membrane Capacitance as Electromechanical Transducer for Pore Gating.

Pacing and clinical electrophysiology : PACE·2015
Same author

An engineering commentary on "Interference of avalanche transceivers".

Pacing and clinical electrophysiology : PACE·2013
Same author

Who should control postmortem explantation?

Pacing and clinical electrophysiology : PACE·2013
Same author

Appropriate parameters for pediatric-specific implantable cardioverter-defibrillators.

Pacing and clinical electrophysiology : PACE·2013
Same author

PACE DATA CARD*

Pacing and clinical electrophysiology : PACE·2013
Same author

To the editor.

Pacing and clinical electrophysiology : PACE·2013
Same journal

Dual Coronary Sinus Lead Strategy to Avoid Tricuspid Valve Traversal in Biventricular Pacing.

Pacing and clinical electrophysiology : PACE·2026
Same journal

A Case of Permanent Pacemaker Implantation via the Epicardial Approach Using the 3830 Lead in an 11-Day-Old Neonate (With Follow-Up of the Above Case).

Pacing and clinical electrophysiology : PACE·2026
Same journal

Cryoballoon Versus Radiofrequency Ablation for Persistent Atrial Fibrillation: Meta-Analysis of Randomized Trials.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Tilt Test Duration in Suspected Vasovagal Syncope: Temporal Patterns and Diagnostic Yield in Patients From Central China.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Combined Leadless Pacing and Subcutaneous ICD Therapy in Long QT Syndromes.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Association of Anesthesia Modality With Procedural Parameters and Clinical Outcomes in PVI for Atrial Fibrillation.

Pacing and clinical electrophysiology : PACE·2026
See all related articles

Professional threshold measurement rules ensure reproducible results in electrostimulation. Key guidelines cover pulse duration, voltage adjustments, and data analysis for accurate rheobase and chronaxie determination.

Area of Science:

  • Electrophysiology
  • Biomedical Engineering
  • Medical Device Technology

Background:

  • Accurate threshold measurement is crucial for effective electrostimulation therapies.
  • Variability in measurement techniques can lead to inconsistent and unreliable results.
  • Standardized protocols are needed for professional threshold determination.

Purpose of the Study:

  • To establish a set of rules for professional measurement of electrostimulation thresholds.
  • To ensure reproducible and comparable results for parameters like rheobase and chronaxie.
  • To provide guidelines for both pacing and defibrillation threshold measurements.

Main Methods:

  • Derivation and discussion of ten specific rules for threshold measurement.
  • Emphasis on voltage averaging, optimal pulse duration, and variable voltage settings.

Related Experiment Videos

  • Recommendations for data analysis, including linear regression and threshold ratios.
  • Main Results:

    • Defined rules for voltage averaging, pulse duration limits (e.g., 1.4 ms for decaying pulses), and step-size selection.
    • Proposed using the average of the last effective and first ineffective pulse for accuracy.
    • Outlined methods for determining strength-duration curves and comparing systems.

    Conclusions:

    • Adherence to the presented rules ensures professional and standardized threshold measurements.
    • The rules facilitate accurate determination of rheobase and chronaxie, even with devices using discrete programming steps.
    • Standardization improves the reliability and comparability of electrostimulation threshold data.